Scale

Setting
Use the Scale screen to:
View current patient weight
Hold a current weight value in memory while equipment is added or removed
Recalibrate scale to zero
Delay patient weighing for a specified length of time
Adjust scale to a known patient weight
Save / access the Weight Trend Chart
Scales / patient weights are for reference only. Scale readings should not be relied upon for
medication dosage.
Unit temperature may affect scale calibration. To ensure accuracy of scale calibration, do not
immediately zero scales when delivered from extreme temperature environments.
1. With the Home screen showing, press the Scale button. The Scale screen will appear with patient's
current weight displayed.
2. Use the buttons on the Scale screen to access any of the Scale functions described as follows.

Hold Patient Weight

Use Hold Patient Weight to retain the current weight value while additional weight (equipment, etc.) is
added or removed from patient surface. Added / removed weight will not be reflected in weight reading
once the Hold Patient Weight button is released.