ALARM CONDITION
Power Loss / Shutdown
At least three people recommended for
this task.
1. Instruct person 2 and 3 to securely hold frame
to prevent shifting.
2. Person 1 disengage emergency release handle.
3. Person 1 pull out lock pin. If lock pin remains
stuck or jammed, have person 2 and 3 gently
shift frame side to side until lock pin is removed.
4. Person 1 re-engage emergency release handle.
5. Follow instructions on screen display to return to
therapy or emergency manual rotation to prone
instructions.
Call ArjoHuntleigh at 1-800-343-0974.
WHAT IT MEANS
Loss of power to the
RotoProne Unit.
IN CASE OF STUCK OR JAMMED LOCK PIN
3
2-26
CORRECTIVE ACTION
Check if power cord has been
disconnected at the head of bed
or at the wall plug. Reconnect,
if necessary.
Do not touch screen until power
cord has been checked. If unit
shutdown is desired, follow the
instructions on the screen.
Lock Pin
Emergency
Release Handle
2
1