Prone / supine settings

Use the Prone and Supine Therapy Settings screen to select and adjust left, center and right pause times
and left and right rotation angles.
Therapy settings should be determined per physician's orders.
1. Select the desired setting to change by pressing the corresponding button location; button will turn
yellow when selected.
2. Press the Adjust up or down arrow button to reach the desired setting.
3. Press the Done button to save all settings and return to the Therapy Settings Selection screen.
4. Press the Home button to return to the Home screen.

Therapy settings Definitions

Patient Left Pause Time: The amount of time the patient is held in place once the target left rotation
angle is reached. Minimum pause time is one minute; maximum pause time is 30 minutes.
Patient Center Pause Time: The amount of time the patient is held in a level position after rotating to the
right or left. Minimum pause time is zero; maximum pause time is 30 minutes.
Patient Right Pause Time: The amount of time the patient is held in place once the target right rotation
angle is reached. Minimum pause time is one minute; maximum pause time is 30 minutes.
Patient Left Maximum Angle: The target left rotation angle to which the patient will turn during rotation
and left pause.
Patient Right Maximum Angle: The target right rotation angle to which the patient will turn during
rotation and right pause.
2-8