Remote mount design

1 Frontend board fixing screw
2 Terminals
3 Frontend board
Figure 69: Replacing the frontend board (flowmeter sensor)
Change from one to two columns
NOTICE
If the O-ring gasket is seated incorrectly or damaged, this
may have an adverse effect on the housing protection class.
Follow the instructions in Opening and closing the housing
on page 38 to open and close the housing safely.
The frontend board can be replaced in the event of a
malfunction.
Component
Frontend board (FEB)
Replace the frontend board as follows:
1. Switch off the power supply.
2. Unscrew / remove the cover.
3. Loosen the fixing screws (3×) at the frontend board.
4. Remove the faulty frontend board.
5. Pull the connector out of the sensor cable harness. Ensure
that the cable harness is not damaged.
6. Pull out the SensorMemory.
4 Connections for flowmeter sensor
5 SensorMemory
Note
The SensorMemory is assigned to the sensor. Ensure that the
SensorMemory remains with the sensor and cannot be lost!
7. Insert the SensorMemory into the new frontend board.
8. Connect the plug of the sensor cable harness.
9. Insert the new frontend board and secure it with the fixing
screws (3×).
10. After powering up the power supply, the transmitter
automatically replicates the system data from the
Ordering number
SensorMemory.
3KXF002564U0100
143