CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E
Change from one to two columns
... 9 Operation
... FillMass batch function

Setup

For the configuration of the fill mass function, the following
steps must be performed:
1. The FillMass function must be active. See also the 'Device
Setup / ...Transmitter / ...Feature Settings / ...' menu.
2. One digital output must be configured as a binary output
with the function 'Batch End Contact'. See also the
'Input / Output / ...' menu. As an option, one digital input
(plug-in card) can be configured with the function
'Batchflow On / Off' at the start of the filling process.
3. The parameters for the fill mass function must be
configured. See also the 'Totalizer / ...FillMass / ...' menu.
Notices on configuration
Damping
During fast filling processes, the damping should be set to
the minimum value to ensure the greatest possible accuracy
of the fill quantity. See also the 'Device Setup / ...Transmitter
/ ...' menu.
Delay until the valve is opened
The delay between the start pulse for the filling process and
the activation of the binary output for opening the valve
depends on the following factors:
From the delay set for the digital input (see parameter
Delay Time on page 111)
From the device internal processing time of 200 ms
Total delay = 'Delay Time' + 200 ms.