CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E
Error code / error text
C078.003
Flowrate to zero
C076.005
All totalizer stopp.
C074.006
Totalizer reset. Reset of one or more
Totalizer
C072.002
Simulation is on. Simulating
process/output value.
C070.026
An alarm is simulated.
S065.028
CO 31/32 readbackcurrent deviates
S060.034
Sensor driver current to high.
S059.035
Density too low.Empty pipe, gas
S058.038
Density to 1g/cm³
S057.031
Sensor temperature out max range
S054.042
Sensor amplitude out of ranges
S052.016
Curr.Out 31/32 is saturated.
S051.017
Curr.Out V1/V2, V3/V4 saturated
Description
External switch-off active via digital input.
Check status of digital input.
Check parameterization.
External switch-off active via digital input.
Check status of digital input.
Check parameterization.
Reset of one or more counters.
Check status of digital input.
Check parameterization.
Simulation mode is active.
Deactivate simulation mode in the 'Diagnostics / ...Simulation Mode' menu.
The alarm simulation is active.
Deactivate alarm simulation in the 'Diagnostics / ...Alarm Simulation' menu.
Current loop error current output 31 / 32.
Check current loop current output 31 / 32 (short-circuit / interruption).
Check load current output 31 / 32.
Check the power supply (voltage) of the transmitter.
Gas bubbles in the measuring tube.
Reduce gas content in the measuring medium.
Deactivate error message by setting the parameter 'Driver Output Max' in the 'Process Alarm /
...Alarm Limits' menu to '0'.
Empty measuring tube. Gas bubbles in the measuring tube.
Reduce gas content in the measuring medium.
Make sure that the meter tube is always completely full.
Deactivate error message by setting the parameter 'Density Low Check' in the 'Process Alarm /
...Alarm Limits' menu to '0'.
Density has been set to 1 g/cm
3
by the transmitter due to an error message.
Contact ABB Service.
Ambient or measuring medium temperature is too high.
Check ambient or measuring medium temperature.
The sensor amplitude is below or above the parameterized limit values 'Sensor Signal Min' and
'Sensor Signal Time'.
Multi-phases measuring medium. Viscosity of the measuring medium is too high.
Check the settings of the parameters in the 'Process Alarm / ...Alarm Limits' menu and adjust if
necessary.
Current output 31 / 32 overshot. The flow has exceeded the set upper measuring range value.
Check the upper range value Qv Max, Qm Max in the 'Device Setup / ...Sensor' menu and correct if
necessary.
Current output V1 / V2, V3 / V4 (plug-in card) overshot. The flow has exceeded the set upper
measuring range value.
Check the upper range value Qv Max, Qm Max in the 'Device Setup / ...Sensor' menu and correct if
necessary.
135
NAMUR classification
Functional check
Functional check
Functional check
Functional check
Functional check
Out of specification
Out of specification
Out of specification
Out of specification
Out of specification
Out of specification
Out of specification
Out of specification