CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E
... 9 Operation

Measurement of standard volumes

Coriolis mass flowmeters can only measure the mass flow of gaseous measuring media.
The operating density of gases is too low to be measured. Consequently the flowmeter is also unable to measure the operating
volume.
However, an appropriate standard volume can be calculated by entering a fixed density for the measuring medium.
The transmitter uses the measured mass flow and the standard density entered to calculate the standard volume flow of the
measuring medium.
(standard volume = mass / standard density).
Configuration of the standard volume measurement using the LCD indicator
Perform the following steps to activate the calculation of standard volumes for gases:
Menu / parameter
1.
... / Device Setup / ...Transmitter / Density Mode
2.
... / Device Setup / ...Transmitter / Density Fixed Value
3.
... / Device Setup / ...Transmitter / ...Units / Unit
Volumeflow Qv
... / Device Setup / ...Transmitter / ...Units / Unit Vol.
Totalizer
... / Device Setup / ...Transmitter / ...Units / Density
Note
To differentiate between the "Fixed density" and "Standard volume calculation" operating modes, we recommend using a customer unit for the volume flow and
volume totalizer units.
4.
... / Device Setup / ...Transmitter / ...Custom Units /
Volumeflow Qv Name
... / Device Setup / ...Transmitter / ...Custom Units /
Volumeflow Qv Factor
... / Device Setup / ...Transmitter / ...Custom Units /
Volume Tot. Name
... / Device Setup / ...Transmitter / ...Custom Units /
Volume Tot. Factor
5.
... / Device Setup / ...Sensor / Qv Max
6.
... / Input / Output / ...Curr.Out 31 / 32 / Uco / Output
Value
7.
... / Display / ...Operator Pages / ...Operator Page 1 / 1st
Line
8.
... / Process Alarm / ...Alarm Limits / Density Max
Setting
Description
Density Fixed Value
Set the 'Density Mode' parameter to 'Density Fixed Value'.
e. g. 1.293 g/l for air
Set the parameter 'Density Fixed Value' to the standard density of
the measuring medium.
e. g. xx/yy for customer unit Set the parameter 'Unit Volumeflow Qv' to the desired unit for
volume flow.
e. g. xx/yy for customer unit, see step 4.
e. g. xx/yy for customer unit Set the parameter 'Unit Vol. Totalizer' to the desired unit for the
volume totalizer.
e. g. xx/yy for customer unit, see step 4.
e.g. g/l
Set the parameter 'Unit Vol. Totalizer' to the desired density unit.
e. g. 'm3/h(qn)'
Set the parameter 'Volumeflow Qv Name' to the desired names for
the standard volume flow. Maximum 8 characters!
e. g. '3.6' with reference to
Set the parameter 'Volumeflow Qv Factor' to the desired factor for
l/s
the standard volume flow. The factor refers to the unit of
liter/second (l/s)
e. g. 'm3(qn)'
Set the parameter 'Volume Tot. Name' to the desired names for the
unit of the standard volume totalizer. Maximum 8 characters!
e. g. '0.001' with reference
Set the parameter 'Volume Tot. Factor' to the desired factor for the
to l
unit of the standard volume totalizer. The factor refers to the unit of
liter (l)
e. g. '1000 m3/h(qn)'
Set the parameter 'Volume Tot. Factor' to the desired upper range
value.
e.g. 'Volume Flow [%]'
Allocate the process variables to the desired outputs of the
transmitter. See Menu: Input / Output on page 107 for further
information.
e.g. 'Volume Flow [unit]'
Allocate the process variables to the operator screens of the LCD
display of the transmitter. See Menu: Display on page 106 for
further information.
Limit value < than
Set the parameter 'Density Max' to a value which is below the
parameter 'Density Fixed
'Density Fixed Value' parameter. This will prevent the 'Density
Value'
exceeds min/max limits.' error message from appearing.