Parameterization Via The Infrared Service Port Adapter; Parameterization Via Hart - ABB CoriolisMaster FCH400 Series Operating Instructions Manual

Coriolis mass flowmeter
Hide thumbs Also See for CoriolisMaster FCH400 Series:
Table of Contents

Advertisement

CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E

Parameterization via the infrared service port adapter

Configuration via the infrared service port adapter on the device
requires a PC / notebook and the FZA100 infrared service port
adapter.
By combining the HART-DTM and the software 'ABB AssetVision'
available at www.abb.com/flow, all parameters can also be set
without a HART connection.
1 Infrared service port adapter
2 USB-interface cable
Figure 60: Infrared service port adapter on the transmitter (example)
1. Position the infrared service port adapter on the front plate
of the transmitter as shown
2. Insert USB interface cable into a free USB female connector
on the PC / notebook.
3. Switch on the device power supply.
4. Start ABB AssetVision and perform the parameterization of
the equipment.
Detailed information on operating the software is available in
the relevant operating instructions and the DTM online help.
Change from two to one column
3 PC / Notebook running ABB
AssetVision and HART DTM
Parameterization via HART®
Configuration via the HART interface of the device requires a
PC / Notebook and a suited HART® Modem.
All parameters can also be set via the HART protocol, using the
HART DTM available at www.abb.com/flow and the ABB
AssetVision software.
1 PC / Notebook running ABB
AssetVision and HART DTM
2 HART modem
Figure 61: HART Modem on the transmitter (example)
For more detailed information on operating the software and
the HART modem, please refer to the relevant operating
instructions and the DTM online help.