CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E
Error code / error text
M038.09
Sensor memory defective.
M037.014
NV chips defect on Motherboard.
M032.022
Curr.Out 31/32 not calibrated.
M031.023
Curr.Out V1/V2 not calibrated.
M030.024
Curr.Out V3/V4 not calibrated.
M028.007
Display value is <1600h at Qmax.
M026.004
Maintenance interval is reached
M024.008
Device not calibrated.
M020.027
Communicat. card not responding
Change from one to two columns
Description
SensorMemory in frontend board defective.
Check if the SensorMemory is defective.
Contact ABB Service.
SensorMemory in motherboard faulty.
Check if the SensorMemory is defective.
Contact ABB Service.
Current output 31 / 32, Uco not calibrated.
Contact ABB Service.
Current output (plug-in card) V1 / V2 or V3 / V4 not calibrated.
Check plug-in card and replace if necessary.
Contact ABB Service.
The current counter reading has exceeded the display resolution.
Check the unit setting for the mass counters / volume totalizers and adjust if necessary.
Maintenance interval reached.
Perform maintenance work.
Start new maintenance interval in the 'Diagnostics / ...Diagnosis Control' menu.
Contact ABB Service.
Fieldbus plug-in is not reacting.
Plug-in card is defective.
Contact ABB Service.
137
NAMUR classification
Maintenance required
Maintenance required
Maintenance required
Maintenance required
Maintenance required
Maintenance required
Maintenance required
Maintenance required
Maintenance required