CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E
Error code / error text
M038.09
Sensor memory defective.
M037.014
NV chips defect on Motherboard.
M032.022
Curr.Out 31/32 not calibrated.
M031.023
Curr.Out V1/V2 not calibrated.
M030.024
Curr.Out V3/V4 not calibrated.
M028.007
Display value is <1600h at Qmax.
M026.004
Maintenance interval is reached
M024.008
Device not calibrated.
M020.027
Communicat. card not responding
Change from one to two columns
Description
SensorMemory in frontend board defective.
•
Check if the SensorMemory is defective.
•
Contact ABB Service.
SensorMemory in motherboard faulty.
•
Check if the SensorMemory is defective.
•
Contact ABB Service.
Current output 31 / 32, Uco not calibrated.
•
Contact ABB Service.
Current output (plug-in card) V1 / V2 or V3 / V4 not calibrated.
•
Check plug-in card and replace if necessary.
•
Contact ABB Service.
The current counter reading has exceeded the display resolution.
•
Check the unit setting for the mass counters / volume totalizers and adjust if necessary.
Maintenance interval reached.
•
Perform maintenance work.
•
Start new maintenance interval in the 'Diagnostics / ...Diagnosis Control' menu.
Contact ABB Service.
Fieldbus plug-in is not reacting.
•
Plug-in card is defective.
•
Contact ABB Service.
137
NAMUR classification
Maintenance required
Maintenance required
Maintenance required
Maintenance required
Maintenance required
Maintenance required
Maintenance required
Maintenance required
Maintenance required