CI/FCM2000-EN CoriolisMaster FCM2000 Commissioning Instruction Mass Flowmeter CoriolisMaster FCM2000 English ® P R O F I PROCESS FIELD BUS B U S...
Data entry ..............................19 Entering data in short form ........................... 21 Appendix ................................22 Other applicable documents ........................22 Approvals and certifications ......................... 22 2 - EN CoriolisMaster FCM2000 CI/FCM2000-EN...
Ex safety information is an integral part of this manual. As a result, it is crucial that the installation guidelines and connection values it lists are also observed. The icon on the name plate indicates the following: CI/FCM2000-EN CoriolisMaster FCM2000 EN - 3...
Approval ABB Automation Products GmbH must be sought for any activities beyond this scope. Repairs performed by ABB-authorized specialist shops are excluded from this. Technical limit values The meter has been designed for use exclusively within the values stated on the name plate and within the technical limit values specified on the data sheets.
Prior to using corrosive and abrasive measurement media, the operator must check the level of resistance of all parts coming into contact with the wetted parts. ABB Automation Products GmbH will gladly support you in selecting the materials, but cannot accept any liability in doing The operators must strictly observe the applicable national regulations with regards to installation, function tests, repairs, and maintenance of electrical products.
Wear to the flange gasket or process connection gaskets (e.g., aseptic threaded pipe connections, Tri-Clamp, etc.) may enable a pressurized medium to escape. When using internal flat gaskets, these can become embrittled by CIP/SIP processes. 6 - EN CoriolisMaster FCM2000 CI/FCM2000-EN...
Type ME24/25 A, U ... MS26 B, E 2. Remote mount design (small nominal diameters) Transmitter − Standard/non-Ex − Ex Zone 2/21, 22 G00387 Flowmeter sensor − Ex Zone 1/21 Fig. 1: FCM2000 overview CI/FCM2000-EN CoriolisMaster FCM2000 EN - 7...
General information on transport Observe the following when transporting the device to the measurement site: • The center of gravity is off center. • Flanged units may not be lifted by the converter housing or terminal box. 8 - EN CoriolisMaster FCM2000 CI/FCM2000-EN...
87 / 88 Sensor 2 5 Gray 6 Violet 7 White 8 Black 9 Orange 10 Yellow 11 "PA" equipotential bonding. When connecting transmitter to flowmeter sensor MS26, transmitter also has to be connected to "PA". CI/FCM2000-EN CoriolisMaster FCM2000 EN - 9...
An additional document with Ex safety instructions is available for measuring systems that are used in explosion hazardous areas (Applies to FM / CSA only). As a result, it is crucial that the specifications and data it lists are also observed. 10 - EN CoriolisMaster FCM2000 CI/FCM2000-EN...
Terminals 33/34 Pulse output Active or passive Terminals 51/52 Contact output Passive Terminals 41/42 Contact input Passive Terminals 81/82 All inputs and outputs are electrically isolated from each other and from the supply power. CI/FCM2000-EN CoriolisMaster FCM2000 EN - 11...
(PE) and the equipotential bonding (PA) in the potentially explosive area. 4.2.1 ATEX/IECEx Ex approval EC type-examination certificate in accordance with ATEX and IECEx KEMA ATEX 08ATEX0150 X, KEMA 08 ATEX 0151X, or IECEx KEM 08.0034X 12 - EN CoriolisMaster FCM2000 CI/FCM2000-EN...
Page 15
When using the device in explosion hazardous areas, the additional temperature specifications in the section titled "Ex relevant specifications" on the data sheet or in the the separate Ex safety instructions (SM/FCM2000/FM/CSA) must be observed. CI/FCM2000-EN CoriolisMaster FCM2000 EN - 13...
If the flow direction indicated on the display is incorrect, it could mean that the signal lead connections have been accidentally reversed. • The locations and the sizes of the fuses can be found in the section titled "Replaceable parts list" in the operating instructions. 14 - EN CoriolisMaster FCM2000 CI/FCM2000-EN...
If the signal outputs are to be connected to intrinsically safe circuits, it is recommended that the included light blue caps be used for the corresponding cable connectors. CI/FCM2000-EN CoriolisMaster FCM2000 EN - 15...
Special information for use in Category 1: • The inside of the meter tube or nominal sizes ≥ DN 50 (2") may correspond to Category 1 (Zone 0). The corrosion resistance of the materials must be taken into account. 16 - EN CoriolisMaster FCM2000 CI/FCM2000-EN...
Visual inspection, particularly of the electronic circuit boards • Visual inspection: no damage or explosion can be detected Zone 1: • Visual inspection: no damage to the threads Ex d, non- (cover, 1/2" NPT cable glands) intrinsically safe circuits CI/FCM2000-EN CoriolisMaster FCM2000 EN - 17...
Page 20
The ambient temperature limits must be observed. If direct sunlight cannot be avoided, appropriate sun protection equipment must be installed. For FNICO or FISCO installations, the number of devices must be limited as per the applicable standard. 18 - EN CoriolisMaster FCM2000 CI/FCM2000-EN...
Access the parameter to be changed and set the new, selected, or default parameter. The ENTER function is effective for approx. 10 s only. If a new value is not entered within 10 s, the transmitter display reverts to the old value. CI/FCM2000-EN CoriolisMaster FCM2000 EN - 19...
Page 22
There are two different methods of entering data: • Numeric entry • Entry from predefined table Important When entering data, the values are checked for plausibility and, if necessary, rejected with an appropriate message. 20 - EN CoriolisMaster FCM2000 CI/FCM2000-EN...
254.50 kg/h Exit "QmMax" STEP or DATA "Progr. level" or nominal size Specialist ENTER "Progr. level" blocked Starting point for "Process information" C/CE → V 98.14 % (transmitter remains online) →V 12,000 m CI/FCM2000-EN CoriolisMaster FCM2000 EN - 21...
ATEX Directive (marking in addition to CE marking) IEC standards Approvals for Canada and United States IMPORTANT (NOTE) All documentation, declarations of conformity and certificates are available in ABB's download area. www.abb.com/flow 22 - EN CoriolisMaster FCM2000 CI/FCM2000-EN...