10 Diagnosis / Error Messages; Calling Up The Error Description; General - ABB CoriolisMaster FCH400 Series Operating Instructions Manual

Coriolis mass flowmeter
Hide thumbs Also See for CoriolisMaster FCH400 Series:
Table of Contents

Advertisement

CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E

10 Diagnosis / error messages

Calling up the error description

Additional details about the error that has occurred can be called
up on the information level.
Process display
Electronics
1. Use
to switch to the information level (Operator Menu).
Operator Menu
Diagnostics
Back
Select
2. Use
/
to select the submenu 'Diagnostics'.
3. Confirm the selection with
Electronics
- F123.321 -
Brief description
Information
Back
The error message is shown on the display according to priority.
The first line shows the area in which the error has occurred.
The second line shows the unique error number. It is made up of
the priority (Fxxx) and the error position (.xxx)
The next lines show a brief description of the error and
information on how to remedy it.
You absolutely need to scroll the display further to read the error
message in more detail.
Note
For a detailed description of the error messages and
information on troubleshooting, see the following pages.
Change from two to one column
.
Exit

General

Errors encountered are itemized in tabular form on the following
pages. The response of the transmitter on error detection is
described therein.
The table lists all possible errors together with a description of
their impact on the value of measurement variables, the
properties of current outputs and the alarm output.
If no entry is indicated in the table field, there is no effect on the
measurement variable or no alarm signal for the particular
output. The sequence of the errors in the table corresponds to
the error priorities.
The first entry has the highest priority and the last has the
lowest.
If multiple errors are detected simultaneously, the error with the
highest priority determines the alarm condition of the
measurement variable and the current output. If an error with a
higher priority does not affect the measurement variable or the
output status, the error with the next highest priority
determines the status of the measurement variable and the
output.