CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E
AssetVision Basic with the ABB CoriolisMaster Device Type
Manager (DTM)
The required software and drivers can be downloaded
using the adjacent download link.
Installation of the software and connection to the flowmeter:
1. Unpack the downloaded archive file to the c:\temp folder.
2. AssetVision Basic (DAT200) install
'3KXD151200S0050_Tool_DAT200_Asset_Vision_Basic'.
3. HART Communication DTM install
'CWCommDTMHART_1.0.55'.
4. CoriolisMaster DTM FCXxxx install
"3KXF410100S0002_DTM_FCXxxx_HART_CoriolisMaster'.
5. Connect the flowmeter with the PC / laptop, see chapter
Parameterization via the infrared service port adapter on
page 69 or Parameterization via HART® on page 69.
6. Power-up the power supply for the flowmeter and start
AssetVision Basic on the PC / laptop
Select HART and 'HART Communication Version 1.0.52'.
Select 'Extended HART modem'.
Select the corresponding COM port.
Activate the 'Multimaster and Burst mode support'
option.
The flowmeter is detected and the CoriolisMaster DTM
starts automatically.
Confirm the dialog field 'Upload parameters' by selecting
'yes'.
COM settings
Baud rate
Number of Stop bits
RTS Control
DTR Control
Parity
Master
Preamble
Number of communication retries
Start Address
End Address
Communication timeout
67
19200
1
Toggle
Enable
Odd
Primary Master
5
3
0
0
10 s