136
CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E
... 10 Diagnosis / error messages
... Error messages
Error code / error text
S049.019
Option Card 1 com error.
S048.020
Option Card 2 com error.
S047.0015
Pulse output is cut off.
S046.000
Mass flowrate exceeds limits.
S044.001
Volume flowrate exceeds limits.
S043.036
Density exceeds min/max limits.
S042.037
Medium temperat exceeds limits.
S041.039
Concentration inunit exceeds
S040.040
Concentration in percent exceeds
Description
Communication error to plug-in card.
•
Check that the plug-in card is installed correctly.
•
Replace plug-in card if necessary.
•
Contact ABB Service.
The pulse rate or the frequency at the pulse output is outside the permissible limits.
•
Check configuration of the parameters for the pulse output.
The mass flow is below or above the configured limit values 'Qm Massflow Min' and 'Qm Massflow
Max'.
•
Check the settings of the parameters in the 'Process Alarm / ...Alarm Limits' menu and adjust if
necessary.
•
Check mass flow.
The volume flow is below or above the configured limit values 'Qv Volumeflow Min' and 'Qv
Volumeflow Max'.
•
Check the settings of the parameters in the 'Process Alarm / ...Alarm Limits' menu and adjust if
necessary.
•
Check volume flow rate.
The density is below or above the configured limit values 'Density Min' and 'Density Max'.
•
Check the settings of the parameters in the 'Process Alarm / ...Alarm Limits' menu and adjust if
necessary.
•
Check density.
The measuring medium temperature is below or above the configured limit values 'Temperature Min'
and 'Temperature Max'.
•
Check the settings of the parameters in the 'Process Alarm / ...Alarm Limits' menu and adjust if
necessary.
•
Check measuring medium temperature.
The concentration in units is below or above the configured limit values 'Concentrat. [u] Min' and
'Concentrat. [u] Max'.
•
Check the settings of the parameters in the 'Process Alarm / ...Alarm Limits' menu and adjust if
necessary.
•
Check concentration.
The concentration in % is below or above the configured limit values 'Concentrat. [%] Min' and
'Concentrat. [%] Max'.
•
Check the settings of the parameters in the 'Process Alarm / ...Alarm Limits' menu and adjust if
necessary.
•
Check concentration.
NAMUR classification
Out of specification
Out of specification
Out of specification
Out of specification
Out of specification
Out of specification
Out of specification
Out of specification