CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E
... 6 Installation
... General installation conditions

Liquid measuring media

Observe the following points to avoid measuring errors:
The meter tubes must always be completely filled with
the measuring medium.
The gases dissolved in the measuring medium must not
leak out. To safeguard this, a minimum back pressure of
0.2 bar (2.9 psi) is recommended.
The minimum vapor pressure of the measuring medium
must be maintained when there is negative pressure in
the meter tube or when liquids are gently simmering.
During operation, there must be no phase transitions in
the measuring medium.
Vertical installation
A
1 Supply tank
2 Sensor
3 Piping constriction / orifice
Figure 9: Vertical installation
B
1
2
3
4
5
G11607
4 Turn-off device
5 Filling tank
For vertical installation in a riser, no special measures are
A
required.
For vertical installation in a downpipe, a piping constriction
B
or an orifice must be installed below the sensor. Doing this
prevents the sensor from draining during the measurement.
Horizontal installation
A
B
Figure 10: Horizontal installation
For liquid measuring media and horizontal installation, the
A
transmitter and terminal box must point upward. If a self-
draining installation is required, the sensor must be mounted
at an incline of ≥ 30°.
Installing the sensor at the highest point of the piping leads
B
to an increased number of measuring errors due to the
accumulation of air or the formation of gas bubbles in the
meter tube.
≥ 30°
G11608-01