Note: The probe zero disk (battery cap) is located on the top end cap of the
gauge. The warning screen will be displayed following the selection and
overwrite of a new transducer.
1) Press the OK key to enter the main measurement screen and begin the
manual zero process.
2) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the probe zero disk.
3) With the "zero probe" hot menu cell highlighted by default, press the ENTER
key to display the confirmation screen, followed by pressing the OK key to
perform the zero.
Note: If the "zero probe" cell is not highlighted, press the MEAS or ESC keys
to highlight.
Performing a Probe Zero