Adjusting The Display - DAKOTA ULTRASONICS MMX-7 Operation Manual

Data logging ultrasonic thickness gauge
Hide thumbs Also See for MMX-7:
Table of Contents

Advertisement

1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right, and the ESC key multiple times to tab left,
until the DISP menu is highlighted and displaying the submenu items.
2) Use the UP and DOWN arrow keys to scroll through the sub menu items
until VIEW is highlighted.
3) Use the LEFT and RIGHT arrow keys to scroll the view options. Once the
appropriate view is displayed, press the MEAS key to return to the
measurement screen and begin taking measurements.

7.3 Adjusting the display

This section will cover the procedures for adjusting the viewable range of the 'scan
bar', in terms of thickness. The B-Scan Start (B-ST) and B-Scan Depth (B-DEP)
used for the B-Scan, will also be used for the 'scan bar' in digits view. The scan bar
displays the viewable thickness range using a starting value, and an overall depth.
Therefore, if the measuring range is 0.500" to 1.00", the start will be 0.500 and the
depth 0.500".
The procedures to adjust the Delay (B-Start & B-Dep) are outlined below:
MMX-7 A/B Scan Thickness Gauge
45