Appendix A Specifications; System Specifications; Sample Pipetting Specifications; Reagent Pipetting Specifications - Dynex Agility Automated ELISA System Operator's Manual

Table of Contents

Advertisement

Appendix A Specifications

A.1

System Specifications

Number of Plates
Number of Sample Tubes
Number of Reagents
Number of Pipettes
A.2

Sample Pipetting Specifications

Sample Tip Size
Sample Pipetting Volume
Time to Dispense
Sample Pipetting Precision
Sample Pipetting Accuracy
Dilution Range
Number of Sample Tips
Sample Tube Dimensions
A.3

Reagent Pipetting Specifications

Reagent Tip Size
Reagent Pipetting Volume
Number of Reagent Tips
Reagent Pipetting Precision
Reagent Pipetting Accuracy
Agility System Operator's Manual
Appendix A Specifications
12 (additional packs can be added using
continuous load capacity)
200
TM
Up to 16 SmartKit
Reagent Packs
simultaneously (additional packs can be
added using continuous load capacity)
2 (sample + reagent)
300 µL
10-300 µL
11 minutes (typical) 50 µL of 96 samples
from plate to sample tubes or deep well
plates
≤ 3% CV (at any operating volume above 10
µL)
± 2% of target volume
1 part in 199 one-stage solution
1 part in 39,601 two-stage solution
20 racks of 112 tips (additional racks can be
added using continuous load capacity)
10-17 mm diameter external dimension
40-100 mm depth
1200 µL
20-1200 µL
1 rack of 98 tips (additional racks can be
added using continuous load capacity)
≤ 3% CV at 10 shots at any volume in
operating range (single shot mode)
± 2% of target volume (single shot mode)
145