Thermo Scientific Dionex Aquion RFIC Operator's Manual page 45

Ion chromatography system
Table of Contents

Advertisement

Temperature Compensation
DS6 Heated Conductivity Cell Components
Thermo Scientific
In ion analysis, the effect of temperature variation is minimized by suppressing eluent
conductivity. To further reduce the effect of temperature variation, a heater inside the cell
regulates the temperature. The cell heater temperature can be set to between 30 °C and 55 °C.
The set temperature must be at least 7 °C above the ambient temperature. Setting the cell
temperature to 0 °C turns off the cell heater.
Built-in preset temperature compensation of 1.7% per °C helps minimize changes in the
baseline or in peak heights when the operating temperature is different from the temperature
at which the cell was calibrated.
The
and
CELL IN
CELL OUT
to the suppressor (see
Figure
component panel.
The cell must be installed at the factory or ordered separately and installed on-site by Thermo
Fisher Scientific field service personnel.
fittings on the front cover of the cell are used to connect the cell
5). The remaining cell components are mounted behind the
2
System Component Details
Dionex Aquion RFIC Operator's Manual