Thermo Scientific Dionex Aquion RFIC Operator's Manual page 136

Ion chromatography system
Table of Contents

Advertisement

6
Service
Replacing the Dionex EGC Holder and Degas Assembly
120
Dionex Aquion RFIC Operator's Manual
Figure 80. Dionex CR-TC installed in Dionex EGC holder
7. Feed the Dionex CR-TC electrical cable through the slot on the side of the Dionex EGC
holder (see
Figure
81).
Figure 81. Dionex CR-TC electrical cable
8. Turn the Dionex EGC holder right-side up and set it on top of the Dionex Aquion RFIC.
9. Direct the
WASTE, GAS SEPARATOR
the rear panel. Snap the line onto one of the tubing clips on the rear panel. Connect the
line to the gas separator waste tube assembly installed on the waste container.
10. Feed the following liquid lines from the bottom of the Dionex EGC holder, through the
tubing chase, to the Dionex Aquion RFIC component panel:
TO SRS/AES REGEN OUT
TO INJ VALVE IN - P/ELUENT OUT
TO PUMP/DAMPER
CR-TC electrical cable
line from the top of the Dionex Aquion RFIC to
Thermo Scientific