Thermo Scientific Dionex Aquion RFIC Operator's Manual page 23

Ion chromatography system
Table of Contents

Advertisement

Thermo Scientific
Figure 3.
Dionex Aquion RFIC front panel LEDs
Table 2. Dionex Aquion RFIC front panel LED states
LED label
Load
Inject
Alarm
Module Connected
Acquiring Data
Pump
Suppressor
Column Heater
If on (green)
If flashing
Injection valve is in
Valve error detected
Load position
Injection valve is in
Valve error detected
Inject position
LED has no "on"
Error detected; check the Chromeleon audit
(green) state
trail for the cause
Dionex Aquion
LED does not flash
RFIC is connected
to a Chromeleon
instrument
Sequence or manual
Sequence stopped because an error was
data acquisition is in
detected
progress
Pump is on
High or low pressure limit exceeded (pump is
turned off )
Suppressor is on and
Continuity check failed, or suppressor is over
current is being
the voltage, current, or power limit
applied to it
(suppressor is turned off )
Column heater is at
Column heater is transitioning to new
set temperature
temperature
2
Operating Features
Dionex Aquion RFIC Operator's Manual
Description
7