Thermo Scientific Dionex Aquion RFIC Operator's Manual page 112

Ion chromatography system
Table of Contents

Advertisement

6
Service
Replacing the Dionex EGC
96
Dionex Aquion RFIC Operator's Manual
Figure 51. Dionex EGC inlet and outlet port lines
INLET
3. Set the Dionex EGC in the service area on top of the Dionex Aquion RFIC. Orient the
cartridge with the electrical cable toward the Dionex EGC holder (see
Figure 52. Dionex EGC in the service area
4. Connect the
line to the
EGC IN
5. Locate the yellow 0.5 mL/min, 7 MPa (1000 psi) backpressure coil (P/N 053765) in the
Dionex Aquion RFIC Ship Kit (P/N 064375).
6. Connect one end of the backpressure coil to the
Figure
53). Leave the other end of the coil unconnected.
Note This backpressure coil connection is temporary; it is used only during the
Dionex EGC conditioning procedure (see
OUTLET
EGC
cable
port on the Dionex EGC.
INLET
OUTLET
page
99).
Figure
52).
port on the Dionex EGC (see