Replacing the Dionex EGC Holder and Degas Assembly
118
Dionex Aquion RFIC Operator's Manual
Figure 76. Dionex EGC in service area
8. Disconnect the
EGC IN
cartridge.
9. Lift the Dionex EGC holder out of the top cover and pull the liquid lines through the
tubing chase (see
Figure
Figure 77. Top cover tubing chase
Tubing
chase
If a Dionex CR-TC is installed, go on to
the new Dionex EGC holder"
If a Dionex CR-TC is not installed, go on to
a Dionex CR-TC"
on
page
To remove the Dionex CR-TC and reinstall it in the new Dionex EGC holder
1. Turn the Dionex EGC holder upside down to view the Dionex CR-TC.
2. Grasp the top of the Dionex CR-TC and lift up to disconnect it from the holder.
3. Disconnect the liquid lines from the four ports on the Dionex CR-TC (see
and
lines from the
EGC OUT
77) to remove them from the system.
"To remove the Dionex CR-TC and reinstall it in
on
page
118.
"To install the new Dionex EGC holder without
121.
and
ports on the
INLET
OUTLET
Figure
Thermo Scientific
78).