Figure 55. Properties dialog box: Eluent Generator tab page
4. Select the EGC-1 check box.
5. Enter the serial number of the Dionex EGC.
6. Verify that the cartridge Type is OH. If it is not, check that you entered the serial number
correctly.
7. Click OK.
To condition the Dionex EGC (KOH, LiOH, NaOH, or MSA)
Always condition a new Dionex EGC for 30 minutes before initial use. This requires directing
the backpressure coil connected to the cartridge outlet to waste, and then generating 50 mM
of eluent at 1.00 mL/min for 30 minutes.
1. Set a temporary waste container (for example, a beaker) on top of the Dionex Aquion
RFIC.
2. Direct the yellow backpressure coil from the Dionex EGC
container.
3. Select these operating conditions on the Chromeleon ePanel:
a. Set the pump flow rate to 1.00 mL/min.
b. Set the Dionex EGC concentration to 50 mM.
Replacing the Dionex EGC
port to the waste
OUTLET
Dionex Aquion RFIC Operator's Manual
6
Service
99