Thermo Scientific Dionex Aquion RFIC Operator's Manual page 122

Ion chromatography system
Table of Contents

Advertisement

6
Service
Replacing the Dionex EGC
106
Dionex Aquion RFIC Operator's Manual
6. Verify that the cartridge type is Carb. If it is not, make sure you entered the cartridge
serial number correctly.
7. Click OK.
To plumb the Dionex K
Figure 65
illustrates the liquid flow path through the Dionex K
conditioning procedure with carbonate eluent.
Figure 65. Dionex K
CO
2
3
1. Disconnect the
TO INJ VALVE IN - P/ELUENT OUT
500 Carbonate Mixer (P/N 088467 for Carbonate Mixer Kit, 2 mm; P/N 088468 for
Carbonate Mixer Kit, 4 mm).
2. Locate the yellow 0.5 mL/min, 7 MPa (1000 psi) backpressure coil (P/N 053765) and a
10-32 to 10-32 coupler (P/N 042627) in the Dionex Aquion RFIC Ship Kit
(P/N 064375).
CO
EGC for conditioning
2
3
EGC plumbing for conditioning
CO
EGC during the
2
3
line from the inlet of the Dionex EGC