Service
Replacing the Dionex EGC
104
Dionex Aquion RFIC Operator's Manual
VORSICHT Die Dionex EGC enthält eine korrodierende Base (KOH, LiOH oder
NaOH), eine korrodierende Säure (MSA) oder eine konzentrierte K
Sie daher beim Umgang mit Kartusche eine Schutzbrille und Handschuhe.
7. Slide the Dionex EGC straight down into the holder until it stops. Make sure the
electrical cable remains outside the holder (see
Note To avoid crimping the lines inside the holder, do not twist the Dionex EGC
when sliding it into the holder. The lines are designed to coil neatly inside the holder.
Figure 62. Dionex EGC installed in the holder
8. Remove the plug from the
9. Orient the blue 4-pin electrical cable connector for the Dionex EGC as shown in
Figure
63. Push the cable connector onto the
ring on the cable connector finger-tight to secure it.
Figure 63. Connecting the Dionex EGC electrical cable
Figure
connector on the Dionex Aquion RFIC top cover.
EGC
connector on the top cover. Twist the
EGC
CO
-Lösung. Tragen
2
3
62).