GE H
RAFT
EALTHCARE
D
5394227, R
IRECTION
EVISION
Section 8-3

Software and Functional Checks after Installation/Upgrade Procedure

8-3-1

Functional Check

Table 8-2
Item
Power ON
System Info
Mode Transition
Section 8-4

Settings - Backup and Restore

This section briefly describes Image Setting backup and restore function.
For detailed documentations and warnings, refer to Chapter.16, Section "System/Backup and Restore
Preset Menu" in the Basic User Manual.
8-4-1

General Information

The Backup/Restore function enables the user to:
Copy/Restore the patient archive.
Copy/Restore the system configuration. The Copy/Restore system configuration feature enables
the user to configure several units with identical presets, providing that the units have the same
software version.
8-4-1-1

Supported Media

Depending on the system, you can use either a CD-R, DVD-R, USB Flash Drive, or USB Hard Disk for
system backup/restore. System does NOT support CD-RW/DVD+R/DVD+RW.
For the sake of simplicity, we have used the CD-R in the examples in this section.
8-4-1-2

General Notes

NOTE:
When you restore backup data from the Utility menu, the LOGIQ S8 application usually restarts
automatically when the restoring is complete.
NOTE:
To perform backup and restore procedures, you must login with administrator privileges.
8-4-2

Backup Procedures

Back up patient data AFTER you've archived (via EZBackup/EZMove) images so that the pointers to
the patients images reflect that the images have been moved to removable media and are no longer on
the hard drive.
1.) Insert a media into the drive or USB device into a USB port.
NOTE:
About formatting media, See 'Formatting removable media' on Basic User Manual for more information.
2.) In the patient screen, select the dataflow Local Archive - Int. HD.
3.) On the touch panel, press Utility.
8-8
Section 8-3 - Software and Functional Checks after Installation/Upgrade Procedure
3
DRAFT (N
Test Procedure Reference
4-3-3 Power On/Off
4-3-4 System Information
4-3-7 Mode Transition Checks
17, 2011)
OVEMBER
LOGIQ™ S8