11
Chapter
Calibrating the Battery
Inspecting for Damage
Cleaning and Disinfecting
070-1149-01 B
Maintenance and Service
NOTE: To ensure safety of patients and operators, periodic safety
checks should be performed by a qualified service technician trained
in medical device safety.
Regular battery calibration is an important part of Eclipse Premier maintenance. Refer
to "Using the Battery" on pg. 2-4 for information on when and how to calibrate the
battery.
WARNING: Hazardous voltage. To reduce the risk of electrical
shock, do not attempt to remove the cover under any circumstances.
Refer servicing to a qualified technician.
Before every use, check the power cord, power plug, power connector, and power input
jack for signs of damage.
Contact an authorized service agent immediately if:
The equipment falls from the cart or is subjected to some other
✔
extreme mechanical stress.
Liquid is spilled on the equipment.
✔
The equipment is not functioning properly.
✔
✔
Parts of the enclosure are cracked, missing or deteriorated.
Any connector or cord shows signs of deterioration such as
✔
cracking.
Clean and disinfect the unit any time it is deemed necessary.
The housing
NEVER use ether, benzene or similar solvents.
CLEANING
Gently rub the housing with a clean, damp cloth using a mild household detergent.
DISINFECTION
Spray the housing with INCIDIN or similar product.
Eclipse Operating Instructions
11-1