Chapter 10
Troubleshooting
10-6
Eclipse Operating Instructions
To reduce electrical interference:
Keep the power cord away from the patient and patient cable.
✔
Connect the unit to a properly grounded wall outlet.
✔
Arrange the patient cable leads together, closely following the
✔
body contour.
Check the line filter setting in the SYSTEM SETUP menu. For
✔
more information, see "Configuring SYSTEM SETUP Menu" on
pg. 4-1.
Ensure that Diathermy or X-ray equipment in adjacent rooms is
✔
not operating. Other electrical equipment including electric beds,
televisions, and lighting fixtures may also generate interference
(even when not in use).
Try moving the patient to another place in the room. Sometimes,
✔
electrical wiring in walls and ceilings causes interference.
Operate the Eclipse Premier from battery power.
✔
070-1149-01 B