Chapter 8
Sending and Receiving Records
Figure 8-2
Back Panel with
Adapter Module
8-2
Eclipse Operating Instructions
Wireless Connection via Adapter Module (Optional)
To connect the adaptor module and the wireless adapter, see the instructions provided
with the wireless device.
WARNING: The Eclipse Premier should not be connected to non-
medical equipment unless the non-medical equipment complies with
IEC 601-1. In addition, the enclosure leakage current of non-
medical equipment connected to the Eclipse Premier should not be
allowed to exceed 500µA in fault condition. To maintain a proper
enclosure leakage current, provide additional protective earth
grounding, use an additional isolating transformer or use a floating
power supply.
WARNING: The Eclipse Premier and any non-medical equipment
that it is connected to should not be allowed within the patient's
vicinity, which extends 6 feet (1.83 m) beyond the perimeter of the
bed, table or chair, and 7 1/2 feet (2.29 m) above the floor.
WARNING: NEVER create a direct connection between an
Eclipse Premier and a PYRAMIS computer if the Eclipse Premier is
already connected to a patient.
070-1149-01 B