Summary of Contents for Cardiac Science POWER2HEART AED G3 Automatic 9300A
Page 2
What We Will Do entity other than Cardiac Science unless specifically authorized by If your Cardiac Science product is returned within 30 days of the date it Cardiac Science. was purchased, at the direction of a customer service representative, we •...
Page 3
What We Will Do entity other than Cardiac Science unless specifically authorized by If your Cardiac Science product is returned within 30 days of the date it Cardiac Science. was purchased, at the direction of a customer service representative, we •...
Page 4
TRADEMARK INFORMATION FirstSave, Powerheart, MasterTrak, MDLink, STAR, IntelliSense, RescueReady, RescueLink, RHYTHMx and Survivalink are trademarks and registered trademarks of Cardiac Science, Inc. Microsoft and Windows are registered trademarks of Microsoft Corporation. All other trademarks are the property of their respective owners.
LIMITED WARRANTY The Cardiac Science AED Operation and Service Manual and any and all information contained herein do not constitute any warranty as to the Powerheart AED G3, FirstSave AED G3, Powerheart AED G3 Automatic or any related products in any manner whatsoever. The “Limited Warranty” is shipped with the AED and serves as the sole and exclusive warranty provided by Cardiac Science regarding Cardiac Science AED products.
CAUTION: This alert identifies hazards that may cause minor personal injury, product damage, or property damage. PRODUCT MODELS This manual is for three, unique Cardiac Science AED models. They share a basic set of features and differences are noted throughout the manual. MODEL...
Canada and +45.4438.0539 (outside the U.S. and Canada.) SAFETY ALERT DESCRIPTIONS The following is a list of Cardiac Science AED safety alerts that appear in this section and throughout this manual. You must read, understand, and heed these safety alerts before attempting to operate the AED.
Page 10
CAUTION: Use only Cardiac Science Approved Equipment Using batteries, pads, cables, or optional equipment other than those approved by Cardiac Science may cause the AED to function improperly during a rescue.
Page 18
MDLink Software by the Medical Director on the Powerheart AED only. CONTINUOUS MONITORING Cardiac Science AEDs are supplied with the Continuous Monitoring disabled. This is the factory default setting. The Powerheart AED can monitor the ECG rhythms continuously throughout the rescue including during charge and CPR mode.
INTELLISENSE BATTERY The Cardiac Science IntelliSense battery technology offers you the most advanced battery capabilities available for defibrillators. Cardiac Science IntelliSense batteries contain an integrated memory chip that automatically stores important usage information, enabling the battery to maintain a complete history of its operating life.
Section 3 - Getting Started BATTERY SHELF LIFE The Cardiac Science AED batteries have a shelf life of five years. Shelf life is defined as the length of time a battery can be stored, prior to installation into AED, without degrading its performance.
CAUTION: Use only Cardiac Science Approved Equipment Using batteries, pads, cables, or optional equipment other than those approved by Cardiac Science may cause the AED to function improperly during a rescue. CAUTION: Possible Improper AED Performance Using pads that are damaged or expired may result in improper AED performance.
As Shown,” pull the second pad from the release liner and place it on the lower left chest, below and left of the breast. Note: Cardiac Science’s standard defibrillation pads are non-polarized and can be placed in either position as shown on the pad package. When using pacing/monitoring pads, refer to placement instructions on the pacing/monitoring pad package.
Page 39
CAUTION: Use only Cardiac Science Approved Equipment Using batteries, pads, cables, or optional equipment other than those approved by Cardiac Science may cause the Powerheart AED to function improperly during a rescue. CAUTION: Possible Improper AED Performance Using pads that are damaged or expired may result in improper AED performance.
Red SERVICE indicator (LED) Maintenance by authorized service personnel is required. is lit. Call Cardiac Science Customer Service (see page 4) or your local Cardiac Science distributor. Red PADS indicator (LED) is lit. Connect the pads or replace with a new pair.
Troubleshooting Table presented in this chapter. If you are unable to resolve the problem, contact Cardiac Science Customer Service for repair information at one of the following telephone numbers: (888) 466-8686, +45.4438.0539 outside the U.S. and Canada, or contact your local Cardiac Science distributor. WARNING: Shock Hazard Do not disassemble the AED! Failure to observe this warning can result in personal injury or death.
AED MODELS 9300 The AED has been designed and manufactured to conform to the highest standards of safety and performance including electromagnetic compatibility (EMC). The Cardiac Science AED Models 9300E/9300A/9300C and pads conform to the applicable requirements of the following:...
ISTA Procedure 2A STAR BIPHASIC WAVEFORM The waveform generated by the Cardiac Science AED is a BIPHASIC TRUNCATED EXPONENTIAL waveform that is compliant with ANSI/AAMI DF2 and DF39. The following is a graph of the waveform voltage as a function of time when the AED is connected to a 50 Ohm resistive load.
Page 54
190-260 ENERGY LEVELS AND PATIENT IMPEDANCE The Cardiac Science Biphasic Truncated Exponential (BTE) waveform utilizes variable energy. The actual energy delivered will vary with the patient’s impedance and the device will deliver a shock when impedance is between 25-180 Ohms. Energy will be delivered at three different levels referred to as ultra-low variable energy, low variable energy, and high variable energy as shown in the above waveform tables.
Meets AAMI DF 39 requirement and AHA recommendation of Specificity – all other rhythms of >95% For detailed information contact Cardiac Science for white papers: P/N 112-2013-001: Pediatric Defibrillation Instructions P/N 110-0033-001: RHYTHMx White Paper P/N 400781: STAR Biphasic White Paper...