NOTICE
Daily/continuous checking
The pump tubing is comprised of wear parts and is susceptible to
leaks and damage.
Check the pump tubing daily and/or continuously.
Replace worn and/or damaged pump tubing.
1 Perform a visual inspection of the pump tubing. Note any cracking or
leaking liquid while doing so.
If the pump tubing should exhibit damage, replace it without delay.
Replacing the pump tubing (see chapter 8.4.5, page 78)
Mounting the press clamp
Prerequisites:
The sample robot is disconnected from the power grid.
1
Figure 51
Mounting the press clamp
Insert one side of the press clamp with the lug into the guide rail
provided for this purpose and push it in.
Plug the other side in as well with the lug in the guide rail and
push it in.
Maintenance
■■■■■■■■
77