Replacing The Seal Of The Sample Beaker Lid - Metrohm OMNIS Sample Robot Pick&Place Product Manual

Table of Contents

Advertisement

Replacing the seal of the sample beaker lid

5.3
Replacing the seal of the sample beaker lid
■■■■■■■■
36
The appropriate seals must be placed on sample beaker lids to tightly seal
sample beakers. These seals must not be damaged.
NOTICE
We recommend replacing the seal of the sample beaker lid approx.
every 6 months.
The standard seals were designed for aqueous applications and Karl
Fischer titrations. If more aggressive solvents such as chlorobenzene
or glacial acetic acid are used, we recommend the more resistant
seals.
Table 8
List of sample beakers with seals for sample beaker lids
Art. no.
Designation
6.02710.000
Lid for sample beaker 75 mL (P&P) with
standard seals
6.02710.030
Lid for sample beaker 75 mL (P&P) with
resistant seals
6.02710.010
Lid for sample beaker 120 mL (P&P) with
standard seals
6.02710.040
Lid for sample beaker 120 mL (P&P) with
resistant seals
6.02710.020
Lid for sample beaker 250 mL (P&P) with
standard seals
6.02710.050
Lid for sample beaker 250 mL (P&P) with
resistant seals
NOTICE
The accessories for the respective product version can be obtained
either on the Internet at
regional Metrohm representative.
http://www.metrohm.com
■■■■■■■■■■■■■■■■■■■■■■
Beaker size
75 mL
75 mL
120 mL
120 mL
250 mL
250 mL
or via your