Blood Glucose Tests
3

Using the Accu-Chek Guide Me System

• The Accu-Chek Guide test strips are for testing fresh capillary whole blood.
• Use only Accu-Chek Guide test strips.
• Use the test strip immediately after removing it from the test strip container.
• Do not apply blood or control solution to the test strip before inserting it into the meter. If you applied
blood or control solution before inserting the test strip into the meter, retest with a new test strip. If a
result appears before applying blood or control solution, do not act on that result.
• Close the test strip container tightly immediately after removing a test strip. Moisture can damage the
test strips and produce incorrect results.
• Store the unused test strips in their original container with the cap tightly closed.
• Discard the test strips if they are past the Use By date printed on the test strip container. If the Use By
date is missing or cannot be read, do not use the test strips. Contact the Accu-Chek Customer Care
Service Center at 1-800-858-8072.
• Refer to the test strip package insert for test strip storage and system operating conditions.
• Do not remove test strips from the test strip container and put them into another container, such as a
plastic bag, pocket, purse, wallet, etc.
• Do not reuse test strips. Once control solution or blood has been applied to a test strip, discard it. If a
retest is necessary, use a new test strip.
• Perform a control test every time you open a new test strip box.
WARNING
To prevent inaccurate results:
• DO NOT expose test strips to heat, moisture, or humidity. Temperatures outside the required
range, as well as moisture and humidity, can damage the test strips and lead to inaccurate
results.
• DO NOT bend, cut, or alter the test strips.
• DO NOT get dirt, food, or other material on the test strip.
18
1080_Guide Me_MANUAL_MASTER.indb 18
3/5/2018 1:27:49 PM

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Guide Me and is the answer not in the manual?

Questions and answers

Colleen Fletcher
February 2, 2025

I have replaced batteries according to the instructions, have checked and recked batteries are in correct positions, now, nothing comes up on screen

1 comments:
Mr. Anderson
February 10, 2025

The Accu-Chek Guide Me meter may not be turning on after replacing the batteries due to the following reasons:

1. Incorrect Battery Installation – Ensure that both CR2032 lithium batteries are inserted with the positive side facing up and properly positioned under the black tabs.
2. Loose Battery Door – The battery door must be securely snapped in place to hold the batteries correctly.
3. Old or Faulty Batteries – Make sure both batteries are new, from the same brand, and properly charged.
4. Damaged Battery Contacts – Check for any dirt or damage on the battery contacts inside the meter.
5. Meter Malfunction – If the issue persists, the meter may need to be reset or checked by customer support.

Try reinstalling the batteries correctly and ensuring the battery door is secure. If the problem continues, consider replacing the batteries again or contacting Accu-Chek Customer Care.

This answer is automatically generated

Delain D Horn
February 15, 2025

what is e-6

Doug Milton
March 15, 2025

Does the Accu Chek Guide Me glucose monitor have a reminder alarm system

KRISHNA MOHAN TANDON
April 1, 2025

what is error 9

Ron Tarpey
April 18, 2025

What is ERR code E-9

1 comments:
Mr. Anderson
May 9, 2025

The E-9 error code on the Accu-Chek Guide Me may still appear after changing the batteries. If it does, remove the batteries again, press the power button for at least 2 seconds, then reinsert the batteries.

This answer is automatically generated