8 Troubleshooting

8.1 Warnings

Malfunction
Warning: 1
Waiting time expired
Continue with button
Warning: 2
Set the time
Continue with button
Sensor mode
Warning: 3
Desiccant used up
Threshold almost reached
Continue with button
Counter mode
Warning: 3
Desiccant used up
Please replace
Press <SET>
Warning: 4
Number of treatments
Remaining: xxx
Continue with button
Warning: 5
Number of treatments
Remaining: xxx
Continue with button
Sensor mode
Warning: 6
Desiccant used up
8 Troubleshooting
8.1 Warnings
Note
Warning messages are displayed to advise the user on correct usage; they do not,
however, prevent use of the unit. Depending on the type of warning, the unit will
either return to normal mode automatically or the user will have to press a button
to acknowledge the warning.
Cause
If an airtight seal cannot be created
at the tooth, the ozone generator
cannot start. Treatment is cancelled
after 60 seconds.
Moisture trap has reached the end
of its service life. Treatment is can‐
celled after 60 seconds.
Blockage in cannula, handpiece or
tubing. Faulty O-ring on tubing or
handpiece.
Treatment time invalid.
After switching on/self test:
Average humidity is ≥ the 7% thres‐
hold that was set.
Desiccant is used up
After switching on/self test:
Counter shows that the program‐
med threshold of 600 hours opera‐
ting time or 1,000 treatments has
been exceeded.
Desiccant is used up.
Unit requires servicing: total permis‐
sible treatment time reached. Xxx
indicates the number of treatments
remaining. An error message is dis‐
played once this number is reached.
Unit requires servicing: total permis‐
sible treatment time reached. Xxx
indicates the number of treatments
remaining. An error message is dis‐
played once this number is reached.
The allocated humidity level is not
reached within the specified timef‐
rame.
Remedy
▶ Continue with any button. Rest‐
▶ Replace moisture trap.
See also: 4.1.2 Install moisture trap,
Page 24
▶ Clean handpiece.
See also: 6.1 Cleaning, Page 39
▶ Press any key to continue.
See also: 5.2.8 Treatment, Page
35
▶ Change desiccant.
See also: 7.2.3 Replace desiccant,
Page 48
▶ Confirm change of desiccant
See also: 7.2.2 Sensor/counter mo‐
de, Page 47
▶ Press any key to continue.
▶ Press any key to continue.
▶ Change desiccant.
51/57
art treatment. Attempt to close
the air circuit within 60 seconds.
Replace O-ring.
Amend treatment time, ensuring
it is within the designated thres‐
holds.
Continue with any button.
with SET button, and confirm
"New desiccant" with "Yes".
The counter is reset to zero.
If the change of desiccant is con‐
firmed with "Yes", the message
reappears after the next use.
Continue with any button.