7 Maintenance
7.2 Maintenance

7.2.5 Ozone neutraliser

Each device is equipped with an internal ozone neutraliser that breaks down the
ozone into oxygen and prevents ozone from leaving the device. The patient and
operator are thereby prevented from excess ozone exposure.
The ozone neutraliser is exchanged at each service interval. A warning is generated
for premature consumption. The device needs to be sent in for servicing.

7.2.6 Service intervals

The unit must be sent away once a year for servicing. Service Technicians are re‐
quired to record all servicing activities on a servicing record located on the interior
of the door.
Safety checks (SFC) are completed in accordance with VDE 0751-1 every two years
as part of the standard servicing activities.
50/57