KaVo HealOzone 2130 C Instructions For Use Manual page 34

Dental ozone machine
Table of Contents

Advertisement

Instructions for use HealOzone 2130 C
|
5 Operation
5.1 Functional description
If the measured current moisture is below the assigned moisture level, the device
switches directly to "Treatment" status. For treatment, the ozone generator is turned
on, and the setpoint for the generator is output. If it takes a while to generate ozone,
"Generate ozone" is shown on the display and reported acoustically with a double
buzzer tone. If ozone cannot be generated within 60 seconds, the warning "Ozone
generation not successful" is output. The message can be acknowledged with any
button, and the treatment process can be restarted. If the ozone is successfully
generated, treatment is started. The treatment time is counted backwards in se‐
conds and shown on the display. A buzzer sounds every second during treatment.
Once the treatment time is over, the device switches to rinse status. The end of
treatment is reported by a long buzzer tone.
32/57

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents