Instructions for use HealOzone 2130 C
|
5 Operation

5.3 Endo mode

CAUTION
5.3 Endo mode
Note
For endodontic treatments, always insert a dental dam to keep from harming the
patient. Sterilise the cannulas before first use.
Danger caused by improper use
Potential 'needle stick' injuries as a result of inserting cannula by hand.
▶ Always use the key to insert and remove cannulae.
▶ Ensure that the O-ring is not worn, damaged or deformed. Replace if necessary.
▶ Using the key, push in the cannula as far as it will go.
▶ Make sure the cannula is positioned securely in the handpiece.
▶ Fit silicone cup onto the handpiece head.
See also: 4.2.3 Attach silicone cup, Page 27
After the root canal has been conventionally prepared, you can continue treatment
with the KaVo HealOzone. A sterile cannula is inserted into the root canal. The
cannula is made of a very flexible and yielding material so that it can follow the
course of the root canal. The cannula penetrates as far as possible toward the apex.
Depending on the remaining tooth substance, the silicone cap seals the crown edge
or gingiva as will all other treatments.
Notes on operator error
If the cannula can penetrate into the root canal, clogging can occur.
If ozone is not applied even though the cap is sealed, the cannula may be plugged.
Screen information:
▪ Treatment time
▪ Seal cap/cannula plugged?
(alternately)