Instructions for use HealOzone 2130 C
|
5 Operation
5.2 The treatment procedure
▪ Change language
▪ Continue

5.2.5 Dryer maintenance

There are two different modes for identifying if the desiccant is used up.
Use the PLUS and MINUS buttons to switch back and forth between the two modes.
Sensor mode (factory setting)
▪ Dryer maintenance
▪ Sensor mode
▪ Change mode
▪ Continue
Counter mode
▪ Dryer maintenance
▪ Counter mode
▪ Change mode
▪ Continue

5.2.6 Treatment start

The treatment process has been activated. The unit waits until a vacuum has been
created. The "Seal cup" line starts flashing.
You can press the stop button or the footswitch to stop the process. Treatment time
can be changed at any point up to the start of treatment. If an airtight seal is not
created within 60 seconds, treatment is cancelled. A warning message appears on
screen.
Screen information:
▪ Treatment time
▪ Seal silicone cup

5.2.7 Dry

If the humidity reading is less than or equal to the allocated humidity level, the unit
immediately switches to treatment mode.
Screen information:
▪ Treatment time