Disinfection; Unit Disinfection; Handpiece Disinfection - KaVo HealOzone 2130 C Instructions For Use Manual

Dental ozone machine
Table of Contents

Advertisement

Instructions for use HealOzone 2130 C
6 Preparation methods DIN EN ISO 17664
CAUTION
|

6.2 Disinfection

6.2 Disinfection

6.2.1 Unit disinfection

▶ Use alcoholic disinfectants such as KaVo ELASTOclean 2 or Ecolab Incidin Li‐
quid to disinfect by wiping.

6.2.2 Handpiece disinfection

Malfunction caused by ultrasound cleaning units.
Potential product damage.
▶ Do not place product in ultrasound cleaning units.
Manual external disinfection
▶ Clean the exterior of the handpiece with chemical disinfectants (wipe disinfecti‐
on) corresponding to manufacturer's instructions for use such as with KaVo
ELASTOclean 2 or Ecolab Incidin Liquid.
Interior disinfection by hand
▶ Connect the handpiece and instrument sleeve.
or
▶ Place the adapter on the quick coupling and rinse for 5 seconds with Spiriclens
(Ecolab) or 3 times with 70% isopropanol (0.2 µm sterile filtered) from the infu‐
sion syringe.
or
▶ Fill with Ecolab Sekusept Plus (0.5%) and let lie for 15 minutes in a water bath,
then rinse out for 15 seconds with a water pistol.
Mechanical external and internal disinfection
Using a Miele thermodisinfector, for instance. Use in accordance with
manufacturer's user instructions. Dry immediately afterwards to ensure product re‐
mains fully functional.
42/57