Instructions for use HealOzone 2130 C
6 Preparation methods DIN EN ISO 17664
|
6.1 Cleaning

6.1.3 Clean plugged endodontics cannula

▶ Push a port spike in and out of both ends of the cannula, then blow compressed
air through cannula.

6.1.4 Clean tubing

Note
Do not place the tubing in ultrasound cleaning units – this could cause malfunction.
▶ Disinfect tubing before cleaning. Disinfect again after cleaning if necessary.
Manual cleaning of the exterior
▶ Clean the exterior of the product by brushing it off under flowing water until there
is no more visible soiling.
Interior cleaning by hand
▶ Connect up the handpiece and instrument tubing.
▶ Place handpiece in a drip pan or basin ①.
▶ Fit adapter ③ onto quick coupling ② and rinse for 15 seconds using a water gun
with cold tap water.
or
▶ Attach a standard 50ml infusion syringe and perform ten 50ml rinses.