A 0.5 P.s Timing (C314); Ns Timing (C807, C814); Gain (R827) - Tektronix 2246 1Y Service Manual

Portable oscilloscope
Table of Contents

Advertisement

h.
Set:
X I0 MAG
SEC/DIV
A INTEN
i.
Select Time CURSORS. Press the TIME MEASURE­
MENT button and select K- SEC-H from the menu.
j.
Rotate the K- OR DELAY control to align the cursor to
the second graticule line.
k.
Rotate the-M control so that the reading is 400.0 jxs.
.
ADJUST- Both the K- OR DELAY control and R823
l
so that the cursors are aligned exactly on the second
and tenth graticule line.
m. Set A INTEN to 10 o 'clock.
4. A 0.5 p.s Timing (C 3 14 )
a.
Set:
X I0 MAG
SEC/DIV
B Trigger MODE
B INTEN
b.
Set time mark generator for 0.5 n.s time marks.
c.
ADJUST—A 0.5
.S TIMING (C314) for one time mark
j j
per division over the center 8 divisions.
5. B 0 .5 p.s Timing (C 3 2 9 )
a.
Set:
Horizontal MODE
SEC/DIV (B)
b.
Set time mark generator for 0.5
c.
Rotate the I*- OR DELAY control counterclockwise
until the readout displays "70.000 p.s" .
d.
ADJUST- B 0.5 jxS TIMING (C329) for one time mark
per division over the center 8 divisions.
REV JUN 1991
Adjustm ent Procedure —224 6 1Y and 2 2 4 6 M O D A Service
6. 2 - 5 ns Timing (C 8 0 7 , C 81 4)
a.
Off
50 )j.s
CCW (Off)
b.
c.
d.
e.
Off
0.5 |j.s
RUNS AFTER
10 o 'clock
B
0.5 jxs
time marks.
jjls
d.
Set:
Horizontal MODE
SEC/DIV
X I0 MAG
CH 1
CH 1 COUPLING
Set time mark generator for 5 ns time marks.
ADJUST-2NSTIM ING (C807,C814)fori cycleper
each 2.5 divisions over the center 8 divisions. See
Figure 5-2. Use the vertical transition of the sine
wave instead of the peaks for better accuracy.
INTERACTION - between C807 and C814. Readjust
as necessary to make the timing at 2.5, 5, and 7.5
divisions within ±0 .2 division (1 minor division).
Disconnect time mark generator.
Figure 5 - 2 . 2 - 5 ns Timing.
7. X Gain (R 827)
a.
Set:
X I0 MAG
Horizontal MODE
Vertical MODE
CH 1 VOLTS/DIV
b.
Connect calibration generator STD AMPL output to
CH 1 input via 50 n coaxial cable. Set generator for
STD AMPL. Set Amplitude to 50 mV.
c.
ADJUST—X GAIN (R827) for 5 divisions of horizon­
tal signal.
Disconnect calibration generator.
A
20 ns
On
50 mV
AC
Off
X-Y
CH 2
10 mV
5 -1 3

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2246 2r2246 mod a

Table of Contents