A djustm ent Procedure —224 6 1Y and 224 6 M O D A Service
Equipm ent Required (S ee Table 4 -1 ):
Time-Mark Generator
Calibration Generator
at the back of this manual for locations of test points and adjustments.
INITIAL CONTROL SETTINGS
Set:
Vertical MODE
INPUT COUPLING
VOLTS/DIV
Vertical POSITION
Controls
Horizontal MODE
A/B SELECT
SEC/DIV
Trigger LEVEL
HOLDOFF
SLOPE
Trigger MODE
Trigger SOURCE
Trigger COUPLING
MEASUREMENTS
MENU Displays
A INTEN
READOUT
FOCUS
SCALE ILLUM
PROCEDURE
1.
H orizontal X I Gain (Timing) (R 826)
a.
Set:
Vertical MODE
CH 1 VOLTS/DIV
b.
Connect time mark generator to CH 1 input via 50 f i
BNC coaxial cable and 50 D, BNC termination. Set
generator for 0.1 ms time marks.
5-12
HORIZONTAL
See the
A D JU S T M E N T LO C A TIO N S
c.
d.
CH 1 and CH 2
DC
0.1 V
2. Horizontal X I 0 Gain (Timing) (R 825)
12 o'clock
A
a.
A
0.1 ms
b.
12 o'clock
min (CCW)
c.
- T
AUTO LEVEL
d.
VERT
DC
OFF
OFF
3. R eadout Horizontal Gain (R 823) and M AG
10 o 'clock
Registration (R 809)
12 o 'clock
for well defined
a.
display
fully CCW
b.
c.
d.
CH 1 (CH 2 off)
e.
0.5 V
f.
g.
50 G Coaxial Cable
50 n Termination
section
Position peaks of the time markers to the center hori
zontal graticule line.
ADJUST-HORIZONTAL IX GAIN (R826) for one
time mark per division over the center 8 divisions.
Set X I0 MAG to ON.
Set time mark generator for 10
Position display about center screen.
ADJUST-HORIZONTAL X I0 GAIN (R825) for one
time mark per division over the center 8 divisions.
Set the time mark generator for 0.5 ms time marks.
Position the middle time marker to the center vertical
graticule line using the Horizontal POSITION con
trol.
Set X I0 MAG to Off.
ADJUST-MAG REG (R809) to position the middle
time marker to the center vertical graticule line.
Set X I0 MAG to On.
CHECK—for no horizontal shift in the time marker.
Repeat parts b through f until no shift is noted.
time marks.
.
jj
s
REV JUN 1991