Equipment Required (See Table 4-1)
Tim e-m ark generator
50 f l BNC termination
1. TRACE ROTATION
a.
Set:
READOUT (Intensity)
A INTEN
Vertical MODE
CH 1 VOLTS/DIV
CH 1 COUPLING
A/B SELECT
Trigger MODE
Trigger SOURCE
Trigger CPLG
Trigger SLOPE
Trigger HOLDOFF
Trigger LEVEL
Horizontal MODE
Horizontal POSITION
A SEC/DIV
Measurements
FOCUS
SCOPE BW
b.
Position trace vertically to the center graticule
line.
Performance Check Procedure—2246 1Y and 2246 Mod A Service
DISPLAY
c.
CHECK—trace rotation control range is adequate
to align trace with center graticule line using a
small straight-bladed alignment tool.
d. ADJUST—trace
For a viewable
graticule line.
readout
For a viewable
trace
2. Geometry
CH 1
a.
Connect tim e-m ark generator (TG 501) to CH 1
0.1 V
via a 50
AC
termination.
A Trigger
AUTO LEVEL
b. Set generator for 0.2 p.s time marks.
VERT
DC
c.
Position the bottom of the CH 1 signal below the
-j~ (positive)
bottom graticule line.
Min
d. CHECK—deviation of any vertical line within the
12 o 'clock
center eight horizontal divisions does not exceed
A
0.1 division (half a minor division).
12 o ' clock
2 |xs
e.
Set CH 1 COUPLING to GND.
All off (Dress
CLEAR
f.
Position trace slowly from the bottom graticule
DISPLAY three
line to the top graticule line while making the fol
times)
lowing check.
For best
defined display
g.
CHECK—bowing or tilt of baseline trace doesn't
Off
exceed 0.1 division (half a minor division) within
the eight vertical divisions.
h.
Disconnect test signal from the 2246 1Y or 2246
Mod A.
50 f l BNC coaxial cable
parallel
to
BNC coaxial cable and a 50 f l BNC
center
horizontal
4-5