To Set Source Output Trigger Parameters; To Set Source Wait Time - Keysight Technologies B2900 Series User Manual

Precision source/measure unit
Table of Contents

Advertisement

NOTE

To Set Source Output Trigger Parameters

For details on this function, see
parameters" on page
4-17.
1. Press the View key to display the Single view.
2. Press the Show Trigger assist key to display the Trigger parameters.
3. Move the field pointer to the Trigger field.
4. Press the rotary knob to change the pointer status to EDIT.
5. Press the AUTO, SYNC, TIMER, or MANUAL assist key to set the trigger
type. The pointer status is changed to MOVE.
6. Move the field pointer to the trigger parameter setup field (Count, Delay, Period,
or Trigger in the Source column).
7. Press the rotary knob to change the pointer status to EDIT.
8. Enter the trigger count (Count), trigger delay time (Delay), trigger period
(Period), or trigger source (Trigger).
9. Press the rotary knob or assist key to fix the setting value.
10. Repeat steps 6 to 9 for all parameters.
If you want to set the trigger parameters in detail, select the MANUAL trigger type
and use the Trigger Configuration dialog box opened by pressing the Trigger >
Config function keys. See

To Set Source Wait Time

For details on this function, see
Control dialog box" on page
1. Press the Config > Common > Wait function keys.
The Wait Control dialog box opens.
2. Set each parameter in the Source column.
3. To apply the settings, press Apply.
To apply the settings and close the dialog box, press OK.
To cancel the setting change, press the Cancel/Local key.
Keysight B2900 User's Guide, Edition 5
"Trigger System" on page 6-27
"Trigger key group" on page
"Measurement Time" on page 6-5
4-31.
Front Panel Operations
To Set Source Output
and
"Trigger
4-37.
and
"Wait
5-17