Front Panel Operations

To Execute Limit Test

To Execute Limit Test
This section describes how to set the limit test and how to display the result of the
limit test. For details on this function, see
To perform the limit test, set the composite limit test, individual limit tests, and
source output and measurement conditions, and then press the Trigger key.

"To Set Composite Limit Test"

"To Set Individual Limit Tests"
"To Display Limit Test Result"
To Set Composite Limit Test
For details on the setup parameters, see
page
4-33.
1. Press the Function > Limit Test > Composite function keys.
The Composite Limit Test dialog box opens.
2. In the Ch field, set the channel to which the setting change is applied.
3. In the Limit Test field, set the composite limit test ON (enable) or OFF (disable).
4. In the Mode field, set the operation mode GRADING (grading mode) or
SORTING (sorting mode).
5. In the Auto Clear field, set the automatic clear ON (enable) or OFF (disable) for
the composite limit test result.
6. When the operation mode is GRADING, set the test result output timing
IMMEDIATE (after every test) or END (after the last test) in the Update field.
7. In the Offset Cancel field, set the offset cancel ON (enable) or OFF (disable).
8. In the Offset field, set the offset value used for the offset cancel.
9. When the operation mode is GRADING, set a bit pattern that represents the limit
test pass state (Pass) in the Pass Pattern field.
10. When the operation mode is SORTING, set a bit pattern that represents the limit
test fail state (Fail) in the Fail Pattern field.
11. In the /BUSY field, set the DIO pin number used for the BUSY signal output.
5-28