Keysight Technologies B2900 Series User Manual page 153

Precision source/measure unit
Table of Contents

Advertisement

Start-up
Sound
Language
Auto CLR
Displays the following softkeys for setting the automatic clear
of the timestamp. The present setting is indicated by an asterisk
on the softkey label.
OFF
ON
Displays the System Start-up dialog box which provides the following setup
parameters.
Power-on State
Power-on Program
Displays the following softkeys to enable or disable the beep and sound. The present
setting is indicated by an asterisk on the softkey label.
OFF
Disables the beep and sound.
ON
Enables the beep and sound.
Displays the following softkeys for specifying the B2900 remote control command
set. The present setting is indicated by an asterisk on the softkey label.
Default
Specifies the default command set which supports all B2900
functions.
2400
Specifies the conventional command set designed for using the
existing programs which you created to control existing
instruments, such as Series 2400 by Keithley Instruments, Inc.
Script
Not effective.
The default command set and the conventional command set are available.
Keysight B2900 User's Guide, Edition 5
Disables the automatic clear of the timestamp.
Enables the automatic clear of the timestamp.
Power-on state, RST, RCL0, RCL1, RCL2, RCL3, or RCL4
The power-on state can be selected from the factory default
reset condition (RST) and user conditions RCL0, RCL1,
RCL2, RCL3, and RCL4 which can be defined by using the
*SAV 0, *SAV 1, *SAV 2, *SAV 3, and *SAV 4 commands,
respectively, while the B2900 is in the remote mode.
Power-on program execution ON or OFF
If this parameter is ON, the power-on program is
automatically executed when the B2900 is turned on.
The power-on program can be defined by using the
:PROG:PON:COPY command while the B2900 is in the
remote mode.
Front Panel Reference
System key group
4-53