For V/I, the channel performs measurement by using the present source/measure
condition, and the resistance value is given by V/I calculation.
Sweep parameters
B2900 can be used as a sweep source, and supports the following sweep operations.
The operation can be selected by using the assist keys displayed when the field
pointer is in the EDIT (green) status on the Sweep Parameters field.
LINEAR SINGLE: Sweep from start to stop in a linear incremental step
LINEAR DOUBLE: Sweep from start to stop to start in a linear incremental step
LOG SINGLE: Sweep from start to stop in a logarithmic incremental step
LOG DOUBLE: Sweep from start to stop to start in a log incremental step
LIST: Sweep of values defined in the List sweep setup list. See
setup".
The following setup parameters are available.
Start
Sets the sweep start value.
Stop
Sets the sweep stop value.
Points
Sets the number of sweep steps.
Step
Sets the sweep step value (incremental step value). Not
available for the LOG and LIST sweep operations.
When the field pointer is in the EDIT (green) status on an entry field, the assist keys
are changed to the unit keys as shown below.
pA, nA, A, mA, and A for current sweep
V, mV, and V for voltage sweep
For the range operation of the sweep source, see
The range parameters Source Volt: Spot and Source Amps: Spot are not effective for
the AUTO and BEST range operations of the sweep source. They are effective for
the constant source and FIXED range operation of the sweep source.
Keysight B2900 User's Guide, Edition 5
Front Panel Reference
Display and Assist Keys
"List sweep
"Sweep dialog box" on page
4-29.