Connect the power cord to the IEC 60320 connector on the rear of the instrument. If
the wrong power cord was shipped with your instrument, contact Keysight
Technologies.
The AC input on the back of your instrument is a universal AC input. It accepts
nominal line voltages in the range of 100 to 240 VAC.

Setting the Power Line Frequency

The power line frequency must be set properly for the AC power at your site. Press
the following function keys to set the frequency to 50 Hz or 60 Hz.
For setting to 50 Hz: More > System > PLC > 50 Hz
For setting to 60 Hz: More > System > PLC > 60 Hz

Setting Date and Time

You can set the date and time on the Date and Time dialog box opened by pressing
System > Info. > Date/Time function keys.

Bench Installation

Do not block the air intake at the sides and the exhaust at the rear of the B2900.
Minimum clearances for bench operation are 5 mm along the sides and 100 mm
along the back.
For easier display viewing and measurement terminal access, you can tilt up the
front of the instrument by turning the handle. To adjust the handle, grab the handle
by the sides and pull outward. Then, rotate the handle to the desired position.