Recommendations For Validation - W&H Lexa MN-111 Instructions For Use Manual

Hide thumbs Also See for Lexa MN-111:
Table of Contents

Advertisement

DISTILLED OR DEMINERALIZED WATER
See the Test instructions service book (conductivity : < 10μS/cm,
Water quality
Total Dissolved Solids: < 6.5 ppm)
Average water
0.17 to 0.2 gal/cycle (0.65 to 0.75 litres/cycle)
consumption
Tank volume
Clean water 1.7 gal (6.4 l)
Min. water charge (clean water): 0.29 gal (1.1 l)
Used water 1.77 gal (6.7 l)
*: usable space with standard rack and trays. With optional racks and trays,
see Accessories, spare parts, consumables.

Recommendations for validation

TEST VALIDATION POINTS
Lexa  | Instructions for Use |  MN-111 ENG Rev14 |  03/04/2023 |  © W&H Sterilization Srl
Part
Description
Hottest points
Coldest points
Technical data
93