10 Appendix
10.2 Service routines
Total filtration [mm]
Additional filter strength [mm] 0
Radiation time level offset
84
10.2.3

Adjustment following increased AL filter value

In some countries an increase in the filter value is called for, for this
purpose there is filter kit 63 36 916.
If an increase in the filter value is not called for, the kit is not required.
Following increase of the AL filter value, the radiation times must be
adapted to the new total filtration.
10.2.3.1

Adapting radiation times

The radiation times must be adapted to the new total filtration.
➢ Call the S28 service routine and set the new total filtration so that the
radiation times adjust automatically.
10.2.3.1.1
S28 service routine
Explanation
Setting the overall X-ray filter strength.
The overall X-ray filter strength is used as a parameter for calculating the
target radiation time and the dose-area product.
Operation
By pressing the +/- button, the filter setting can be amended between 1.5
mm AL and 2.5 mm AL in 0.1 mm steps.
A = Basic filter value
B = Changeable filter value
Acknowledge the setting by clicking the Adult key.
By pressing the Child key, the setting will be rejected and the service
routine is exited.
Assigning the filter strength for the radiation time
Using the software, an offset is set for the radiation time according to the
filter strength.
>1.5
>1.6
>1.7
>1.8
0.1
0.2
0.3
0
0
0
+1
Installation Instructions Heliodent Plus
>1.9
>2.0
>2.1
>2.2
0.4
0.5
0.6
0.7
+1
+1
+1
+2
Dentsply Sirona
>2.3
>2.4
>2.5
0.8
0.9
1.0
+2
+2
+2
62 14 972 D 3507
D 3507.031.01.15.02 03.2018